Choose only one dye for each PMT.
Please discuss your choices with flow core personelle
as not all combonations are workable.

Please note the following are not options:
BFP, CFP, TEXAS RED, Mrfp, HOESCHT, FURA-2, INDO-1, INDO-RED
,or any of the ALEXA 500s.

Excitation Laser Detector Filter nm Dye
Blue 488nm FSC Diode 488/10 (483-493) Forward Scatter
PMT A 780/60 (750-810) PE-Cy7
PMT B 695/40 (675-715)
PMT C 576/26 (563-589)
PMT D 530/30 (515-545)
PMT E 488/10 (483-493) Side Scatter
Red 633nm PMT A 780/60 (750-810)
PMT B 660/20 (515-545)
Violet 405nm PMT A 530/30 (515-545)
PMT B 450/40 (430-470)

Click image to enlarge
flow cytotmetry filters

For more info on protocols: Click Here