After connecting circuit, power ON with switch located on back panel. Unit goes through self-test and requires calibration of flow and O2 sensor. Flow and O2 sensor will then require calibration at the same time every day. Flow calibration takes less than 20 seconds.




Check-out tests
Settings
After connecting ventilator to baby and checking that chest rises, go into
List and view Setting values to refine ventilator settings to optimize flow
waveform.
2. Check that there is not a space between end of inspiration and beginning of exhalation using the flow/time graphic
3. Check that there is no air-trapping using flow/time graphic
4. Pressure limiting inspiration allows user to increase PIP Limit setting to increase tidal volume. However, if PIP Limit is increased and peak inspiratory pressure (PIP) does not increase to equal new PIP Limit setting, then increase the set Flowrate until peak inspiratory pressure is being limited again.
5. When pressure limiting ventilation, an increase in the PEEP/CPAP level will decrease the tidal volume if the PIP Limit is not increased by the same amount as well.